Ecotec E3000 Messages 6-1
kina22e chapter 6.fm technical handbook(1206)
5 Ecotec E3000 Messages
During leak detection operations the LCD display will indicate information which
supports the operator in running the Ecotec E3000. Besides measurement data, also
current equipment conditions, operating hints as well as warnings and error
messages can be displayed.
5.1 Error Messages and Warnings
The Ecotec E3000 is equipped with comprehensive self-diagnostic functions. When
a faulty condition is sensed by the control board, this condition is indicated to the
operator via the LCD display as far as possible.
Errors
Errors are events which force an interruption of the measurements in progress and
which the Ecotec E3000 is not capable of rectifying on its own. Errors are indicated
in plain text together with an error number.
The Ecotec E3000 remains in the error status. After the fault cause has been
removed the error can be acknowledged by pressing the
RESTART button. The Ecotec
E3000 then returns to the measurement mode.
Warnings
Warnings are issued if an abnormal condition is detected that may cause increased
inaccuracies of measurements but will not interrupt measuring completely.
Notice: In case of enquiries please keep the serial number and the software
version number of the Ecotec E3000 at hand.
The following abbreviations are used in the listing + error messages below:
In the following table all possible error messages and the recommendations stated
in the display in case of this error are listed. If no recommendation is given, please
call your nearest INFICON service representative (see Section 1.2.1).
TMP turbo molecular pump
TC TMP controller
TSP Transpector NV-RAM: non-volatile RAM