11 - 15
HAPSITE ER Operating Manual
42 Click Delete. (See Figure 11-43.)
Figure 11-43 Deleting Point
NOTE: If a point was inadvertently deleted, the original calibration points can
be restored by clicking the Restore Initial Record button. (See
Figure 11-44.)
Figure 11-44 Restoring Initial Record
43 If a point is missing because it does not meet the peak search criteria, the peak
search parameters can be adjusted. Click Search settings to adjust all of the
compounds at once. To adjust individual analytes, click View/Edit. Refer to ()
for more information. (See Figure 11-45.)
Figure 11-45 Adjusting Peak Search Settings