tira85e1-a (2014-10) PxG55x.cp 17
Safe Value: The value to be used for Safe State = Safe Value.
• PCG gauges with ATM: The "Manufacturer Configuration Module 3" is assigned
to the Pirani module.
Index SI DataType NV Access PM Name
4020 0x01 BYTE RW Safe State
0x02 REAL RW Safe Value
Specifies the behavior for the Value for states other than Valid.
Safe State
0
1
2
3
Zero
Full Scale
Hold last value
Use safe value
Safe Value: The value to be used for Safe State = Safe Value.
Index SI DataType NV Access PM Name
6000 0x0E BOOL RO tx TxPdoState
0x11 REAL RO tx Sensor Value
Is set if the device is not in Safe State (value (I 0x6nn0, SI 0x11) = valid)
TxPdoState
0
1
Invalid
Valid
The corrected, converted, calibrated final analog input value of the sensor.
Index SI DataType NV Access PM Name
6001 0x01 BOOL RO tx Reading Valid
0x02 BOOL RO tx Overrange Exceeded
0x03 BOOL RO tx Underrange Exceeded
Indicates whether the Value parameter contains a valid value within the specified
accuracy or not.
Reading Valid
0 Invalid
1 Valid
Subindex 0x02
4.3.3 Manufacturer
Configuration Module 3
(PCG with ATM only)
Subindex 0x01
Subindex 0x02
4.4 Input Area PCG with
ATM (0x6000…0x6FFF)
4.4.1 Input Common
Subindex 0x0E
Subindex 0x11
4.4.2 Input Capacitance
Diaphragm
Subindex 0x01