Index SI DataType NV Access PM Name
8001 0x01 REAL x RW Zero Adjust Mode
The CDG Sensor is adjusted automatically by the Pirani sensor, if this index is set.
The High Trip Point is calculated by:
High Trip Point = Value
*)
× Percentage
**)
*)
Value from Piezo Sensor (I: 0x6010, SI: 0x11)
**)
Percentage High Trip Source (I: 0x800E/F, SI: 0x13)
In this case, the High Trip Point value is automatically updated by the device itself.
Any user-written value to the parameter High Trip will be refused with an abort
code.
The Low Trip Point is a fixed value:
The value defined in Low Trip Point Limit is compared with the pressure value
referenced by the Source Index parameter.
High Trip Point
Time t
in-
active
activein-
active
active
Low Trip Point
Measurement signal
(Pressure p)
Low Trip
Hysteresis
High Trip
Hysteresis
in-
active
Index SI DataType NV Access PM Name
800E 0x01 BOOL x RW High Trip Enable
0x02 BOOL x RW Low Trip Enable
0x11 REAL x RW High Trip Point Limit
0x12 UDINT x RW High Trip Source Index
0x13 REAL x RW Percentage High Trip
Source
0x14 REAL x RW Low Trip Point Limit
0x15 UDINT x RW Low Trip Source Index
0x16 REAL x RW Percentage Low Trip
Source
0x17 REAL x RW High Trip Hysteresis
0x18 REAL x RW Low Trip Hysteresis
0x1A UDINT x RW Source Index
4.7.2 Configuration
Capacitance Diaphragm
Subindex 0x01
4.7.3 Configuration Trip
Point 1