High Trip Hysteresis: Hysteresis value for High Trip Point.
Low Trip Hysteresis: Hysteresis value for Low Trip Point.
Object index of active source of (I 0x900F SI 0x01) Input Value.
Source Index
Bit 16…31
Bit 08…15
Bit 00…07
Index
Subindex
Reserved
This chapter describes the configuration area for the PSG.
The High Trip Point is a fixed value
The Low Trip Point is a fixed value:
The value defined in Low Trip Point Limit is compared with the pressure value
referenced by the Source Index parameter.
High Trip Point
Time t
in-
active
activein-
active
active
Low Trip Point
Measurement signal
(Pressure p)
Low Trip
Hysteresis
High Trip
Hysteresis
in-
active
Index SI DataType NV Access PM Name
800E 0x01 BOOL x RW High Trip Enable
0x02 BOOL x RW Low Trip Enable
0x11 REAL x RW High Trip Point Limit
0x12 UDINT x RW High Trip Source Index
0x14 REAL x RW Low Trip Point Limit
0x15 UDINT x RW Low Trip Source Index
0x17 REAL x RW High Trip Hysteresis
0x18 REAL x RW Low Trip Hysteresis
0x1A UDINT x RW Source Index
Subindex 0x17
Subindex 0x18
Subindex 0x1A
4.9 PSG Configuration Area
(0x8000…0x8FFF)
4.9.1 Configuration Trip
Point 1