54
tina60e1-a (2011-04)
6.3 Troubleshooting
In case of an error, it may be helpful to just turn off the
mains supply and turn it on again after 5 s.
Problem Possible cause Correction
Output signal per-
manently
≈0V
Sensor cable defective or not
correctly connected
Check the sensor cable
No supply voltage Turn on the power supply
Error Remedy the error
Gauge in an undefined status Turn the gauge off and on
again after 5 s (reset)
FAIL PIR1 Pirani sensor defective Replace the sensor
(
→ 53)
Electronics unit not correctly
mounted on sensor
Check the connections
(electronics – sensor)
FAIL EEPROM EEPROM error Turn the gauge off and on
again after 5 s (reset)
Replace the gauge
FAIL SENSOR Electronics unit not compatible
with the sensor
Replace the sensor
(
→ 53)
Replace the gauge