8.4 Replacing the SL200 sniffer line filter
Structure of the sniffer
tip
Fig.15:
Sniffer tip
1 Capillary filter (plastic; standard) 4 Cross-head screws
2 Felt pads 5 Sinter filter
3 Capillary filter (metal with gasket;
option)
6 Gasket
A blockage of the sniffer tip can have the following causes:
Capillary filter blocked -> replace felt discs or capillary filter
Sinter filter clogged -> check or replace sinter filter
Capillary clogged in sniffer tip -> replace sniffer tip
Sniffer tip damaged -> replace sniffer tip
Sniffer line damaged / blocked -> replace SL 200.
Replace felt discs or
capillary filters
Switch off the leak detector or disconnect the sniffer line from the device.
Use tweezers to remove the dirty felt discs and replace them with new ones.