INFICON Accessories and interfaces | 10
UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109) 131 / 152
ð The modes "Standard" or the mode "Inverse" are available for selection.
To mode "Inverse": In the following table overview, the arrow pointing the
transfer direction in the column "Transition" must be reversed.
5
Save .
Function Transition Description
VENT inactive→ active: Vent inlet (only in standby state)
CAL inactive→ active:
active→ inactive:
Start calibration.
Apply value for background and finish calibration.
No function - The input is not functioning. Recommended setting of unused inputs.
Clear inactive→ active: Erase warning or error message / cancel calibration.
Sniff inactive→ active:
active→ inactive:
Enable sniffer mode.
Enable vacuum mode.
Purge inactive→ active:
active→ inactive:
Switch-on purge (only in standby state)
Switch-off purge (only in standby state)
Start inactive→ active: Change to the measurement mode.
Start/Stop inactive→ active:
active→ inactive:
Change to the measurement mode.
Switch to Standby.
Stop inactive→ active: Switch to Standby.
ZERO inactive→ active:
active→ inactive:
Switch ZERO on.
Switch ZERO off.
10.2.4 Configuring digital outputs
The digital outputs PLC-OUT 1 ... The available functions can be assigned in any way
necessary to the 8 I/O module.
ü
Supervisor
rights
1
> Setup > Accessories > I/O module > Configuration of digital outputs
2
Select the desired "Digital output".
ð The outputs PLC-OUT 1 - 8 are available.
3
Select the desired "Function", see the table below.
4
Select the desired "Mode".
ð The modes "Standard" or the mode "Inverse" are available for selection.
To mode "Inverse": In the table overview the description of the state "open"
must be replaced with the description of the state "closed".
5
Save .