10 | Accessories and interfaces INFICON
134 / 152 UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109)
10.3.1 Creating a connection between the device and the bus
module
To establish the connection between the leak detector and the bus module, proceed
as follows:
1
Switch off the leak detector.
2
Connect the INFICON bus module via a data cable to the LD socket of the
device, see “Connections for accessories and control signals [}31]“.
3
Switch on the leak detector.
4
> Setup > Accessories > Device selection
5
Select the "Bus module" as device.
6
Confirm with .
10.3.2 Setting a bus module address
You can set the setpoint for the field bus address. With PROFIBUS® this is the node
address. DeviceNet is the MAC ID.
This setting is irrelevant for Profinet and EtherNet/IP.
1
> Setup > Accessories > Bus module
2
Enter the desired address in the "Bus module address" field.
3
Confirm with .
ð The value set is first carried over when restarting the leak detector. To do this,
switch the power supply off and back on.