10 | Accessories and interfaces INFICON
136 / 152 UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109)
10.4.1.2 Setting the LAN connection in the PC or Tablet
LAN connection - quick start
If you have performed the steps described here once, it is sufficient to enter the IP
address for many devices in case of repetition.
ü The PC is connected to the same network as the leak detector.
ü A LAN setting has been set up in the configuration of the leak detector, see also
“Configure the LAN connection of the leak detector [}135]“.
ü You have noted down the IP address of the leak detector. It can be found in the
leak detector under "Diagnosis > Information > Device > Network".
ü Java Script is enabled in the web browser. We recommend using a current version
of the web browser Chrome™, Firefox® or Safari®.
►
To access the leak tester from the PC or tablet web browser, enter the IP address
of the leak detector as follows:
http://<IP address>
ð The current active user interface of the leak detector is called.
ð The same functionality is available on the PC or Tablet as on the touchscreen of
the leak detector.
10.4.1.3 Allow client access
ü
Supervisor
rights
1
> Setup > Network > Client access
2
To allow the leak detector to be operated by PC or Tablet, activate the "Client
Access" option.
ð If the option "Client Access" is not activated, the leak detector cannot be
operated via PC or Tablet. No settings can then be changed.
3
Save .
Display of all connected clients
ü The network connection between the leak detector and one or more PCs or Tablets
has been established, see “Setting the LAN connection in the PC or Tablet [}136]”.
► > Setup > Network > Connected clients