EasyManua.ls Logo

Inficon UL3000 Fab ULTRA - Setting the LAN Connection in the PC or Tablet; Allow Client Access

Inficon UL3000 Fab ULTRA
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 | Accessories and interfaces INFICON
136 / 152 UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109)
10.4.1.2 Setting the LAN connection in the PC or Tablet
LAN connection - quick start
If you have performed the steps described here once, it is sufficient to enter the IP
address for many devices in case of repetition.
ü The PC is connected to the same network as the leak detector.
ü A LAN setting has been set up in the configuration of the leak detector, see also
“Configure the LAN connection of the leak detector [}135]“.
ü You have noted down the IP address of the leak detector. It can be found in the
leak detector under "Diagnosis > Information > Device > Network".
ü Java Script is enabled in the web browser. We recommend using a current version
of the web browser Chrome™, Firefox® or Safari®.
To access the leak tester from the PC or tablet web browser, enter the IP address
of the leak detector as follows:
http://<IP address>
ð The current active user interface of the leak detector is called.
ð The same functionality is available on the PC or Tablet as on the touchscreen of
the leak detector.
10.4.1.3 Allow client access
ü
Supervisor
rights
1
> Setup > Network > Client access
2
To allow the leak detector to be operated by PC or Tablet, activate the "Client
Access" option.
ð If the option "Client Access" is not activated, the leak detector cannot be
operated via PC or Tablet. No settings can then be changed.
3
Save .
Display of all connected clients
ü The network connection between the leak detector and one or more PCs or Tablets
has been established, see “Setting the LAN connection in the PC or Tablet [}136]”.
> Setup > Network > Connected clients

Table of Contents