EasyManua.ls Logo

Inficon UL3000 Fab ULTRA - Meaning of the Status LED; Assembly of the Touchscreen

Inficon UL3000 Fab ULTRA
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INFICON Description | 4
UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109) 27 / 152
4.3.2.5 Meaning of the status LED
The status LED is located on the back of the instrument inside of the connector block
for the equipment and control signals, see also "Connections for accessories and
control signals [}31]”.
Status LED Meaning
Flashing red No communication with operating unit
Flashing blue Run up
Blue Standby
Flashing green Evacuate (pumping)
Flashing green Calibrating
Flashing green Software is started (immediately after the
device is switched on)
Flashing green Boot loader is waiting for a software
update (after starting the software update
of the basic unit)
Green Measuring
Yellow Measuring with warning not
acknowledged
Red Error
4.3.2.6 Assembly of the touchscreen
The display is a touchscreen.
The touchscreen responds to being touched lightly. To correctly select the chosen
function, avoid strong pressure.
You can always see symbols for the navigation on the display.
Additionally, you can see, depending on the context, other symbols and elements.

Table of Contents