6 | Operation INFICON
54 / 152 UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109)
5
To assign the required rights to the user, select a group. Via and select
between the groups User, Operator and Supervisor. See "Overview of Rights
Groups [}52]".
6
In the field "Language" assign a language to the user via and .
7
Save .
6.2.3.3 Modify Personal Settings
As a user with limited rights (
User
) you can also modify your language or PIN. By this
the associated user profile is correspondingly adapted. Access to the entire user
profile is not necessary.
1
Press on your name which appears on the top left of the display.
ð The window "User accounts" opens.
2
Select the “Change own PIN” or “Change own language” button as required.
See also
2 Logging off from the device [}93]
6.2.4 Switch off Automatic Login
Factory setting
Login: Supervisor
PIN: 1111
Language: English
As per factory settings, after switching on the device the user "Supervisor"
automatically logs in and the measurement screen is called. This default user also has
the permissions of the group "Supervisor". Without changing this setting, any user can
operate all functions without restriction.
You can specify that the login window appears after you turn on the device instead of
automatically logging on.
In the login window, all users can login, which have been already registered in the
device, see"Select, modify, create user profile [}53]”.
ü
Supervisor
rights
1
> User accounts > Manage Automatic Login
2
In the “Manage Automatic Login” window, deactivate the option “Active”.
3
Save .
ð After restarting the device, the current settings are applied.