6.2.6.1 Changing the presentation of the line graph
Fig.11:
Presentation of the line graph
ü
Operator
or
Supervisor
rights
1
> Display > Line chart
2
In the field "Scaling" select between "Linear" and "Logarithmic".
3
Select between the different viewable ”Decades”.
4
To dynamically adjust the upper and lower limit on the leak rate, activate the
option "Autoscale".
5
In the field "Time axis" select the length of the time axis "30", "60", "90", "120" or
"240" seconds.
6
If the option "Autoscale" is not active, select "Lower diagram limit" in the field the
desired decade.
7
Save .