6.2.6.2 Changing the presentation of the bar graph
Fig.12:
Presentation as bar graph
ü
Operator
or
Supervisor
rights
1
> Display > Bar chart
2
In the field "Scaling" select between "Linear" and "Logarithmic".
3
Select between the different viewable ”Decades”.
4
To dynamically adjust the upper and lower limit on the leak rate, activate the
option "Autoscale".
5
If the option "Autoscale" is not active, select "Lower diagram limit" in the field the
desired decade.
6
Save .