INFICON Operation | 6
UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109) 67 / 152
6.3 Settings for the measurements
6.3.1 Select operation mode
ü
Operator
or
Supervisor
rights
1
> Operation mode
2
Choose between “Vacuum”, “Sniff / SL200”, “Auto Leak Test“ and “Sniff /
SL3000”.
3
Save .
6.3.2 Select gas
DANGER
Danger from a Hydrogen explosion
Hydrogen can explode in combination with oxygen. The allowable composition of
venal gas mixtures can be read in the safety data sheets of the respective
manufacturers.
Ensure that the share of hydrogen does not exceed the described concentration.
ü
Supervisor
rights
1
> Setup > Measurement > Mass
2
Make a selection from:
"Hydrogen" H
2
(2 amu)
"Helium"
4
He (4 amu)
"Mass 3"
3
He (3 amu)
3
Save .
6.3.3 Setting setpoints
You can set the leak rate for the setpoint to 1, 2, 3 and 4 separately.
When the setpoints are exceeded:
If the setpoint 1 or 2 is exceeded, the measurement line in the measurement
window changes color.
The setpoint relay of the digital output switches, see also "I/O module [}128]" or
the interface description.
Also setpoint 1 defines the trigger point for the different alarms, see also "Change
audio settings [}60]". It is also the limit for I•ZERO, see “Set and use the function
ZERO [}72]“.
ü
Operator
or
Supervisor
rights