6 | Operation INFICON
68 / 152 UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109)
1
> Setpoints
2
Adjust.
3
Save .
6.3.4 Activate vacuum ranges
Vacuum ranges
You can activate the vacuum ranges MASSIVE, GROSS, FINE and ULTRA for your
measurements.
All 4 areas can be activated at the same time. If several areas are activated, they
switch automatically depending on the inlet pressure. In this way different sensitivities
can be measured with.
At least one range must be activated.
ü
Operator
or
Supervisor
rights
1
> Setup > Measurement > Vacuum > Vacuum ranges
2
Activate the desired vacuum range under "Vacuum range".
3
Save .
6.3.5 Set Auto Leak Test
Please take note of the information on the operation mode "Auto Leak Test" and the
notes on measuring. See also Operation mode “Auto Leak Test” [}20], “Select
operation mode [}67]”, “Measuring [}80]” and “Result display of the Auto Leak Test
[}59]”.
ü
Operator
or
Supervisor
rights
1
> Setup > Measurement > Vacuum > Auto Leak Test
ð You can edit the following settings:
- “Min. Test time”
- “Max. test time”
- “Leak rate setpoint 1”: To change the setpoint 1. The same threshold value
can alternatively be set in the window "Setpoint values", see “Setting
setpoints [}67]”.
- “Part numbering”: After activating this function, a number is assigned for
each test. This facilitates the traceability of test results, see “Call result log
[}87]”.
- “Part number”: To assign a part number. For each measurement, this partial
number is counted as starting number.
- “Warning limit for consecutive failed tests”: After the entered number has
been reached, a warning message is displayed.
2
Adjust.
3
Save .