6 | Operation INFICON
78 / 152 UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109)
When using hydrogen or helium-3 (
3
He) as tracer gas an internal calibration is not
possible. In this case, use an external calibration leak.
The operation modes "Vacuum" and "Sniffing" must be separately calibrated when
using an external calibration.
A calibration should only take place when the device is at operating temperature.
Calibrate at least 20 minutes after switch-on.
For systems with a greater distance to the test object, we recommend the use of an
external calibration leak at the location of the test object.
When using the internal calibration leak, observe the setting of the machine factor,
see also “Setting the machine factor [}70]“.
6.3.16.1 Internal calibration
ü User rights
ü The device is in the state "Standby" or "Measure".
1
Select the cross-hair icon .
2
Select the calibration “Internal”.
3
Press .
ð The device performs an internal calibration.
6.3.16.2 External calibration
ü
Operator
or
Supervisor
rights
1
Mount the calibration leak on the inlet of the device.
2
> Setup > Measurement > Calibration leak
3
Adjust the leak rate of the calibration leak used, see also “Set external
calibration leak [}71]”.
4
Place the device in the state "Standby" or "Measure".
5
Press the cross-hair icon .
6
Select the calibration “External”.
7
Press .
8
Follow the instructions on the screen.
6.3.16.3 Check calibration
The calibration can also be checked without changing calibration factor. This tells you
whether a new calibration is necessary.
ü User rights
ü The device is in "Standby" or "Measure" mode.