6 | Operation INFICON
84 / 152 UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109)
When accessing via LAN, use the IP address found under > Device >
Network.
When accessing via WLAN, use the URL found under > Setup > Network
> WLAN.
Example 1:
http://192.168.11.124:3000/md?f=csv_en&c=4&d=mr
Returns measured values and metadata of the measurement cycle with ID 4 in the
format “csv_en”.
Example 2:
http://ul3xxx.inficon.com:3000/md?dts=2018-11-05T23:00:00.000Z&dte=
2018-11-06T23:00:00.000Z&f=json&d=r
Returns measured values between 2018-11-05T23:00:00.000Z und
2018-11-06T23:00:00.000Z in the “json” format.
Parameter Name Description Options Example
c Measureme
nt cycle
Querying one or more
measurement cycles via
ID
ID's of the
measurement cycles.
Separated by comma
for several
c=3
c=3, 5, 8
dts START The start time of the time
period from which data is
to be exported.
A date in ISO format
in UTC time
dts=2018-08-02T17:15:12.000Z
dte End The end time of the time
period from which data is
to be exported.
A date in ISO format
in UTC time
dte=2018-08-02T17:45:35.000Z
f Format The data format of the
export
json, csv_en, csv_de
Standard: csv_en
f=json
f=csv_en
d Scope Scope of the data. With
or without metadata
mr: Meta and
measured values
m: Meta only
r: only measured
values
Standard: mr
d=mr
d=r
6.6.4 Measurement data database: Information
The maximum size of the database corresponds approximately to the number of
measured values that are generated over the period of one month during continuous
recording. This corresponds to a database size of approx. 1.5 GB.