INFICON Operation | 6
UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109) 93 / 152
4 Mass spectrometer 9 Fore pump
5 Turbo molecular pump 10 V1 … V10: Solenoid valves for controlling
gas flows
6.13 Resetting to factory settings
You can restore the device to factory settings.
Loss of settings and measurement data
After resetting to factory settings only the manufacturer factory settings are in the
memory of the device. The software of the device is not reset when restoring the
factory settings.
ü
Supervisor
rights
1
> Setup > General > Reset
2
Make your choice:
ð If you want to reset all settings, press the "Reset settings" button.
ð If you want to delete all recorded measurement data and user accounts,
press the "Delete data" button.
ð To reset the device to the factory settings, press both buttons one after the
other.
6.14 Logging off from the device
1
Press your name, which is displayed in the upper left corner of the display, or
select > User accounts.
ð The window "User accounts" opens. See also "Modify Personal Settings
[}54]".
2
You log off from the device via the button "Log off".
ð The login window opens.