EasyManua.ls Logo

Inficon UL6000 Fab - Replacing the SL3000 Sniffer Line Filter

Inficon UL6000 Fab
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 | Cleaning and maintenance INFICON
118 / 152 UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206)
If you have activated the display of maintenance requests, set the maintenance
interval to the desired new period. Also see "Activate or deactivate maintenance
requests [}61] " and "Set maintenance interval "Filter sniffer tip" or "Air filter" [}61]".
8.5 Replacing the SL3000 sniffer line filter
Replacing the filter
inserts of the
capillary filter and the
water conservation
tip
The plastic capillary filter, the metal capillary filter and the water conservation tip are
equipped with filter inserts.
There is a plugwasher below the metal capillary filter and the water conservation tip.
This seal is already integrated into the plastic capillary filter.
Table3: Capillary filter
1 Sniffer line end 4 Plastic capillary filter
2 cone seal 5 Water conservation tip
3 Metal capillary filter
Capillary filter and water conservation tip are screwed onto the sniffer line end. The
filter inserts are inserted and lie in front of a small metal grill that is also inserted.
To replace the filter inserts, proceed as follows:
1 Switch off the leak detector.
2 Unscrew the capillary filter or the water conservation tip. Make sure that the
cone seal does not fall out.
3 Push out the old filter pads and the metal grill from the rear.
Fig.17: Pressing the filter inserts out of the capillary filter
4 Dispose of the old inserts and clean the metal grill.

Table of Contents

Related product manuals