10 | Accessories and interfaces INFICON
130 / 152 UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206)
10.2 I/O module
10.2.1 Create a connection between the device and the I/O
module
ü Supervisor rights
1 Connect the INFICON I/O module with a data cable to the LD socket on the
rear of the device, see "Connections for accessories and control signals [}31]".
2 > Setup > Accessories > Device selection
3 Select the "IO module".
4 Save .
10.2.2 Configuring analog outputs
ü Supervisor rights
1 > Setup > Accessories > I/O module > Configuration of analog outputs
2 Set-up the "Analog output 1".
ð You can choose from "Off", Pressure p1", Pressure p2", "Leak rate mantissa",
"Leak rate exponent", "Leak rate linear", "Leak rate logarithmic", "Leak rate
mantissa hyst.", and "Via interface".
3 Set-up the "Analog output 2".
ð You can choose from "Off", Pressure p1", Pressure p2", "Leak rate mantissa",
"Leak rate exponent", "Leak rate linear", "Leak rate logarithmic", "Leak rate
mantissa hyst.", and "Via interface".
4 Set-up the "Analog output scaling".
ð You can choose from "0.5 V / decade", "1 V / decade", "2 V / decade", "2.5
V / decade", "3 V / decade", "5 V / decade", or "10 V / decade".
5 Set-up the "Exponent upper limit".
6 Save .
The settings "Analog output scaling" and "Upper limit exponent" affect both analog
outputs dependent on the selected function, see the table overview below.
Functions and assignment of analog outputs:
Off The analog outputs are
switched off
(Output voltage = 0 V).