60 tinb29e1-c (2019-10) VGC083A_B.om
Pressure
Voltage
Pressure
Voltage
Pressure
Voltage
Note - An analog output of above +11 volts indicates a faulty convection
gauge or unplugged gauge cable.
The equations and coefficients for the nth order polynomial equation defined for
various pressure measurement ranges for the Non-linear analog output (S-curve)
for CG1 or CG2 for N
2
/air described above are given in the following table:
For 0 mTorr to about 2 Torr, the Non-Linear Analog Output voltage range of
0.375 to 2.842 volts, use this table.
Coefficients for y(x) = a + bx + cx
2
+ dx
3
+ ex
4
+fx
5
For 2 Torr to about 100 Torr, the Non-Linear Analog Output voltage range of
2.842 to 4.945 volts, use this table.
Coefficients for
(
)
=
++
2
2
3
For 100 Torr to 1000 Torr, the Non-Linear Analog Output voltage range of 4.94 to
5.659 volts, use this table.
Coefficients for
(
)
=
+
2
Where y(x) = pressure in Torr, x= measured analog output in volts
Example: Measured analog output voltage is 0.3840 V.
From first table shown above use equation:
y(x) = a + bx + cx
2
+ dx
3
+ ex
4
+fx
5
x = 0.3840 volts
a = -0.02585, b=0.03767, c=0.04563, d=0.1151, e=-0.04158, f=0.008738
y(x) = Pressure = 1.0E-03 Torr