tina96e1-b (2016-12) VGC501_VGC502_VGC503.om 77
CAL corresponds to the COR command
Transmit: CFx [,a.aaa] <CR>[<LF>]
Description
x Calibration factor gauge x =
1 –> Gauge 1
2 –> Gauge 2
3 –> Gauge 3
a.aaa Calibration factor gauge x, 0.100 … 10.000
(default = 1.000)
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a.aaa <CR><LF>
Description
a.aaa Calibration factor gauge x
Transmit: COR [,a.aaa,b.bbb,c.ccc] <CR>[<LF>]
Description
a.aaa Calibration factor gauge 1, 0.100 … 10.000
(default = 1.000)
b.bbb Calibration factor gauge 2
c.ccc Calibration factor gauge 3
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a.aaa,b.bbb,c.ccc <CR><LF>
Description
a.aaa Calibration factor gauge 1
b.bbb Calibration factor gauge 2
c.ccc Calibration factor gauge 3
5.6 Gauge Parameters
CAL - Calibration factor
CF1, CF2, CF3 - Calibration
factor gauge 1, 2 or 3
COR - Calibration factor