Inficon VGC502 - 5.7 Gauge Control; SC1, SC2, SC3 - Gauge 1, 2 or 3 control

Inficon VGC502
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
84 tina96e1-b (2016-12) VGC501_VGC502_VGC503.om
Transmit: SCx [,a,b,c.ccEscc,d.ddEsdd] <CR>[<LF>]
Description
x Controlled gauge, x =
1 –> Gauge 1
2 –> Gauge 2
3 –> Gauge 3
a Gauge activation, a =
0 –> Manual (default)
1 –> Hot start
3 –> Via measurement channel 1
4 –> Via measurement channel 2
5 –> Via measurement channel 3
b Gauge deactivation, b =
0 –> Manual (default)
1 –> Self control
3 –> Via measurement channel 1
4 –> Via measurement channel 2
5 –> Via measurement channel 3
c.ccEscc ON threshold (s = sign)
d.ddEsdd OFF threshold (s = sign)
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c.ccEscc,d.ddEsdd <CR><LF>
Description
a Gauge activation
b Gauge deactivation
c.ccEscc ON threshold (s = sign)
d.ddEsdd OFF threshold (s = sign)
5.7 Gauge Control
SC1, SC2, SC3 - Gauge 1, 2 or
3 control