6 | Operation INFICON
66 / 108 XL3000flex-Operating-instructions-jina83en1-07-(2105)
2
If you want to perform a calibration, insert the sniffer tip into the calibration
opening again and then immediately press the right button on the sniffer line.
ð Otherwise the calibration would only be checked again.
Change settings
The aforementioned behavior can be changed by configuring the "CalMate mode" in
the menu "Main menu > Settings > Set up > Accessories > CalMate". You have a
choice of 3 options there:
0 =
PROOF / CAL
Inserting the sniffer tip into the calibration opening starts the calibration check.
Pressing the right-hand button of the sniffer line switches to calibration (factory
setting).
1 =
CAL only
Inserting the sniffer tip into the calibration opening starts the calibration. It is not
possible to check the calibration.
2 =
PROOF only
Inserting the sniffer tip into the calibration opening starts the calibration check. It is not
possible to switch to calibration.
6.3.4.7 Entering the calibration factor
The calibration is usually determined by the appropriate calibration routine. Therefore,
it is usually not necessary to adjust the calibration factor manually.
An incorrectly set calibration inevitably leads to wrong leak rate indicator!
6.3.4.8 Calibration factor sniffing
Entry of the calibration factors for masses
2, 3, 4 in low flow and in high flow.
The values will be overwritten during the
next calibration.
"High Flow-" or XL settings are available
only in operation mode "XL Sniffer
Adapter".
The calibration factors are managed
separately to earth and to "High Flow"
and "Low Flow".
0.01 … 100
Control unit Main Menu > Settings > Set up > Operation modes > SNIF >
Calibration factor