Item No. 8105097 12/2013 Page 3 of 4
8. Measure the REAR and match target between 1kHz and 4kHz.
o Select the appropriate FM-specific target:
FM Boom or
FM Chest
the choice should reflect what is used with the FM system being tested
o Measure the REAR by clicking the Start button.
o Adjust volume control on the FM receiver to match the targets between 1kHz and 4kHz.
9. Re-measure the RESR/MPO at user setting.