Immunitytest
Test/Compliancelevel
Electromagneticenvironment–guidance
Voltagedips,short
interruptionsand
voltagevariations
onpowersupply
inputlines
IEC61000-4-11
<0%U
T
for0,5cycleat
45°steps
0%U
T
for1cycles
70%U
T
for25/30cycles
<5%U
T
for250/300
cycles
Mainspowerqualityshouldbethatofatypicalcommercialorhospital
environment.Iftheuserofthisproductrequirescontinuedoperationduring
powermainsinterruptions,itisrecommendedthattheproductispowered
fromanun-interruptiblepowersupplyorabattery.
U
T
isthea.c.mainsvoltagepriortoapplicationofthetestlevel.
Powerfrequency
(50/60Hz)
magneticeld
IEC61000-4-8
30A/m
Powerfrequencymagneticeldsshouldbeatlevelscharacteristicofatypical
locationinatypicalcommercialorhospitalenvironment.
ConductedRF
IEC61000-4-6
3V
150kHzto80Mhz
6V
inISM&amateurradio
bands
Fieldstrengthsfromxedtransmitters,suchasbasestationsforradio
(cellular/cordless)telephonesandlandmobileradios,amateurradio,AM
andFMradiobroadcastandTVbroadcastcannotbepredictedtheoretically
withaccuracy.ToassesstheelectromagneticenvironmentduetoxedRF
transmitters,anelectromagneticsitesurveyshouldbeconsidered.Ifthe
measuredeldstrengthinthelocationinwhichthisproductisusedexceeds
theapplicableRFcompliancelevelabove,thisproductshouldbeobserved
toverifynormaloperation.Ifabnormalperformanceisobserved,additional
measuresmaybenecessary,suchasreorientingorrelocatingthisproduct.
Interferencemayoccurinthevicinityofequipmentmarkedwiththefollowing
symbol:
281574470-D