= 200 kg
8. Technical specifications. Technical specifications
Dimensions (HxBxD) 19x47x25 cm.
Max. lifting capacity 200 kg.
Number of lifts per charge 65 lifts of 0.5 m with 80 kg.
Lifting velocity at 200 kg load 2 cm/s.
Max lifting interval 250 cm.
Height from celling to sling hook 41 cm.
Total weight without sling 7 kg.
Electrical connection of charger 230V AC 50Hz
Voltage/current between charger/hoist 24V DC 900mA / 50mA.
Charging time (with 2 Ah battery) Max 2 hours
Battery 20 pcs. 1.2V / 2.0 Ah 4/5 SC NiMH
Intermittence 8%
Sound pressure 50-55 dB (A)
Protection class
IPX4
8.1 Symbols
Warning.
SWL (Safe Working Load) = Max. load (patient + sling + other accessories).
Class II equipment.
Type B equipment - patient not separated from ground.
CE - marking.
The product should be reused where possible.