EasyManua.ls Logo

inventum TMO430 - Page 42

inventum TMO430
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Immunity test IEC 60601
Test level
Compliance
level
Electromagnetic
environment - guidance
Electrical fast
transient/ burst
IEC 61000-4-4
�2 kV for
power
supply
lines
�1 kV for
input/
output
lines
Not
applicable
The ME equipment is
solely battery powered.
Surge IEC
61000-4-5
�1 kV
differential
mode
�2 kV
common
mode
Not
applicable
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3A/m Complies Power frequency
magnetic fields should
be at levels characteristic
of a typical location in a
typical commercial or
hospital environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
> 95% dip
0.5 cycle
60% dip
5 cycles
70% dip
25 cycles
95 dip 5
sec.
Not
applicable
The ME equipment is
solely battery powered.
42
English

Related product manuals