10.2 Test Whole Blood Controls
Procedure:
10.2.1 Insert the test cuvette with the blood reservoir facing up.
10.2.2 When the messages Add Sample and Press Start are displayed, dispense
one drop of whole blood quality control material into the center sample well
of the test cuvette.
Note: Refer to the HEMOCHRON
®
Jr. Signature microcoagulation instrument
Operator’s Manual for details on filling the test cuvette.
10.2.3 Press the Start button. After the test, read the results from the display.
10.2.4 Repeat Steps 1 through 3 two more times, using a new test cuvette for each
replicate.
10.2.5 Examine the results that were obtained. If no fault messages were displayed
and if the results on the replicates are within the Acceptable Range labeling
for the whole blood quality control material, the system performance is
acceptable.
If fault messages were displayed, refer to the Fault Messages section of this
manual.
If results on the replicates are not within the Acceptable Range labeling for the
whole blood quality control material, verify the system temperature as outlined in the
HEMOCHRON
®
Jr. Signature microcoagulation instrument Operator’s Manual. If
results are still unacceptable, contact ITC for assistance.
HEMOCHRON
®
Jr. Signature Service Manual 14