Parameter setup and measuring operation
LDM52 OEM, Release 003, Revision 000
Value selection:
nn
ASCII
Meaning
1
0x0D0A
CR LF
2
0x0D
CR
3
0x0A
LF
4
0x02
STX
5
0x03
ETX
6
0x09
Htab (Tabulator)
7
0x20
Space
8
0x2C
Single Quote
9
0x3A
Colon
10
0x3B
Semicolon
When an invalid character is entered, it will not be set. The current separator will be kept
instead.
6.4.19 TP - Internal temperature
Output of the internal device temperature in °C
The internal temperature is about 10 kelvins higher than the ambient temperature.
When the internal temperature of the laser unit is 75°C the laser will be switched off.
Warning w1906 will be generated in a cycle of 10 sec.
After the cooling of LDM52 OEM and reaching of an internal temperature of +70°C the inter-
rupted measurement starts automatically (depends on the setup with DT or CT).
The LDM52 OEM will not be switched off automatically when the temperature is below the in-
ternal temperature of -10°C.
Query:
TP
Response (example): 26°C
Please see the notes in chapter 3.1.