10.10.4 CIP/SIP definition (JUMO digiLine pH only)
Open CIP/SIP Definition Digital Sensors:
Device Menu > Configuration > Digital Sensors > Digital Sensors 1 to 6 > CIP/
SIP Definition
10.10.5 Calibration timer
Open Calibration Timer Settings Digital Sensors:
Device Menu > Configuration > Digital Sensors > Digital Sensors 1 to 6 > Cali-
bration Timer
Calibration timers prompt the user to calibrate sensors on a regular basis. The
settings for all analysis inputs and the universal inputs are explained together.
⇨ Chapter 10.12 "Calibration timer", page 174
Configuration point Selection/
setting option
Explanation
CIP min. tempera-
ture
a
-20 to+150 °C Temperature thresholds for identifying CIP/SIP
cycles
If the CIP/SIP cycle takes place above one of
these values during the set duration of a CIP/SIP
cycle, the values identify a successfully complet-
ed CIP/SIP cycle and the CIP or SIP counter in
the JUMO digiLine electronics is incremented.
The respective counter is reset only after the val-
ue has dropped below the CIP/SIP temperature.
SIP min. temperature
a
CIP cycle duration
a
0 to 9999 s Duration of a CIP/SIP cycle
SIP cycle duration
a
CIP/SIP alarming inactive
active
Setting for alarming on the JUMO AQUIS touch
P, if the CIP, SIP or autoclaving counters in the
JUMO digiLine electronics have reached the
maximum number of permissible cycles.
a
This setting is saved in the configuration of the JUMO digiLine electronics.