10 Configuration
186
10.16 Washtimer
Open: Device Menu > Configuration > Washtimer > Washtimer 1 to 2
Configuration point Selection/
setting option
Explanation
Washtimer active Yes
No
Activation/deactivation of the washtimer
Washtimers are used to clean analysis sensors
regularly and can actuate external equipment for
cleaning sensors via binary outputs.
Designation up to 20 text characters Timer designation
Time interval 0 to 999 h Cycle for automatically repeated sensor cleaning
Washing time 0 to 999 min Duration of sensor cleaning
Hold time after
washing
0 to 999 s After-run time of the washtimer hold signal upon
timeout of the washing time
The washtimer hold signal provides an On-signal
for the duration of the washing time plus the
hold time.
The hold signal is intended primarily to activate
the hold mode for controllers and analog outputs.
These functions are
held in a safe mode until the washed analysis
sensor provides stable measured values once
again.
Signal for
interval restart
Yes
No
Activation/deactivation of event list entries during
sensor cleaning
Event list text up to 21 text characters Text for the event list during sensor cleaning