16 Calibrating CR conductivity sensors
232
16.3.2 Calibrating the temperature coefficient
Step Action
1 Start calibration of the temperature coefficient.
Device menu
r Calibration r Select CR Analysis Input r TC Cal-
ibration
2 Clean the sensor and immerse it in the sample solution. Ensure
that the rel. cell constant is calibrated correctly (if necessary, make
a trial measurement with a test solution).
For CR optional boards, continue with step 3;
for universal inputs set to the "conductivity measurement" opera-
tion mode, continue with step 4
3 Enter one of the measuring ranges 1 to 4. Confirm the entry by
pressing "OK"
The calibration values determined apply only to the selected mea-
suring range.
tap button to enter
measuring field