17 Calibrating Ci conductivity sensors
238
Calibration default settings for calibrating the TC curve
Parameter Setting options Explanation
Temperature com-
pensation
Selection from
analog selection
Temperature input for automatic sensing of the test/
sample solution temperature during the calibration
Starting temperature -50 to 250 °C The starting and final temperatures of the range for
which a temperature coefficient curve is to be deter-
mined.
The starting temperature must be at least 20 °C low-
er than the final temperature. The reference tem-
perature for the measuring input must lie between
the starting and final temperatures, and differ from
the starting and final temperatures by at least 2 °C.
Final temperature -50 to 250 °C
NOTE!
Calibration of the TC curve is possible only with automatic temperature sens-
ing.