239
17 Calibrating Ci conductivity sensors
17.3 Ci calibration routines
17.3.1 Calibrating the relative cell constant
NOTE!
Conductivity measuring inputs can be configured with measuring range change-
over. Accordingly, calibrations must be performed for all "accessible measuring
ranges".
NOTE!
You must be logged in with corresponding user rights to perform calibrations.
v Chapter 8.2.1 "Log-on/Log-out", Page 89
NOTE!
Analysis inputs for inductive conductivity measurements (Ci) must undergo a
basic Ci calibration in the course of commissioning. Subsequent calibration is
not possible without this initial basic Ci calibration.
v Chapter 11.2 "Ci base calibration", Page 200
Step Action
1 Start calibration of the relative cell constant.
Device Settings Menu
r Calibration r Select Ci Analysis Input or
Universal Input
r Relative Cell Constant Calibration
for Ci optional boards, continue with step 2;
for universal inputs set to the "conductivity measurement" operat-
ing mode, continue with step 3;
2 Enter one of the measuring ranges 1 to 4. Confirm the entry by
pressing "OK"
The calibration values determined apply only to the selected mea-
suring range.
tap button to enter
measuring range