17 Calibrating Ci conductivity sensors
240
3 Ensure that
• the sensor has been cleaned and is immersed in the test solu-
tion,
• the set reference conductivity matches the conductivity value of
the test solution.
Wait until the measuring value displayed stabilizes and then con-
firm the result of the measurement by pressing "OK".
The preset reference conductivity can be changed here manually
if necessary.
4 A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
5 Press "Yes" to accept the calibration values determined and enter
the calibration in the calibration logbook.
Press "No" to discard the results.
Step Action
tap button to change
reference range