EasyManua.ls Logo

JUMO AQUIS touch P - Page 259

JUMO AQUIS touch P
404 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
257
19 Calibrating O-DO sensors
O-DO calibration default settings
The calibration default settings enable the calibration routines to be accessed in
the particular calibration menu.
Calibration routines that are not enabled are not visible in the calibration menu.
Additional calibration default settings are explained in the following table.
Parameters Possible settings Explanation
Compensation Fixed compensation temperature
Temperature input
Interface
Fixed compensation temperature:
Compensation with a fixed temperature
value that is entered in the configuration
item "Compensation temperature".
Temperature input: The integrated
temperature probe of the O-DO sensor
supplies the compensation tempera-
ture.
Interface: The AQUIS touch p transmits
the compensation temperature to the
sensor electronics via the interface. The
source of the compensation tempera-
ture is set in the configuration point
"Compensation temperature".
Temperature com-
pensation
Selection from Analog selection Temperature input for automatic sens-
ing of the test/sample solution tempera-
ture during the calibration

Table of Contents

Related product manuals