259
19 Calibrating O-DO sensors
4 The device displays the current measured value of oxygen satura-
tion and the test material temperature. The setpoint value for oxy-
gen saturation (100 %Sat) is set automatically by the device. The
user does not need to make any entries here.
5 Wait until the measured value displayed stabilizes and then con-
firm the result of the measurement by pressing "OK"
6 A protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
7 Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.
Step Action