21 Calibration during disinfection measurements
21.1 Important information
21.2 General information
JUMO digital sensors for disinfection measurands (chlorine, hydrogen peroxide,
peracetic acid, ozone, chlorine dioxide and bromide) are calibrated by means of
suitable comparison measurements in the measured value for the sensor type
concerned (e.g. DPD method). Be sure to observe the information in the oper-
ating manual of the sensor for disinfection measurands with respect to calibra-
tion. The concentration values from the comparison measurement are entered
in the device during the calibration process.
WARNING!
During the calibration, the relays and analog output signals assume the config-
ured states! The response of the output signals is set for each output in its "Re-
sponse at calibration" configuration point.
Chapter 10.5 "Analog inputs", page 139