22 PC Setup Program
330
22.8.10 Buffer set tables
Three buffer set tables are created in the JUMO AQUIS touch P.
Buffer set tables are needed for calibration of pH sensors when using automatic
buffer detection.
Chapter 13 "Calibrating a pH measuring chain", page 211
.
These Buffer set tables can be changed as often as necessary here.
The factory settings for the Buffer set tables are as follows:
Buffer set 1: Reference buffer solutions for calibrating pH measuring instru-
ments acc. to DIN 19266
Buffer set 2: Technical buffer solutions, preferably for calibrating and adjust-
ing technical pH measuring instruments acc. to DIN 19267
Buffer set 3: No data
Open Settings:
Only Setup > Buffer Set Tables > "Buffer Set Table 1 to 3“ tab