22 PC Setup Program
354
22.9 Diagnostics
22.9.1 Calibration logbooks
The calibration logbooks for all analysis inputs can be displayed on both the
device and under this menu item in the JUMO PC setup program.
Chapter 12.3 "Calibration logbook", page 201
22.9.2 Digital sensor calibration logbooks
The calibration logbooks for all digital sensors can be displayed on both the
device and under this menu item in the JUMO PC setup program.
Chapter 12.3 "Calibration logbook", page 201
22.9.3 Digital sensor connection list
If the JUMO AQUIS touch P must be replaced, e.g. because of a defect in the
device, and a backup of the configuration is available as a setup file, the config-
uration can be loaded into the new device. However, the original links of the
digital sensors to the corresponding inputs for the digital sensors must be rees-
tablished manually. The information required to do this can be displayed with
the PC setup program. The setup file backup from the original device contains
the list of link data. Open this setup file and click in the navigation area of the
PC setup program:
Diagnosis > Digital Sensor Connection List
The link data is displayed in this list. By comparing this data, the sensors can
be reassigned to their original inputs and linked manually.
Chapter 8.2.7 "Digital sensors", page 101