355
22 PC Setup Program
Procedure for linking digital sensors following device replacement:
Step Action
1 Using the JUMO PC setup program, load the originally backed up
configuration into the device.
2 Press the "Print" button or File > Print
3
In the print selection list, select "Digital Sensor Connection List".
4 Print out the list (confirm the selection with "OK")
5 On the device, open the menu Device Menu > Digital Sensors,
scan, and link all connected sensors.
Chapter 8.2.7 "Digital sensors", page 101