67
7 Startup
7.1 Initial startup
WARNING!
Prior to startup, ensure that the device has been installed and connected prop-
erly and
in compliance with the installation instructions. Observe the safety instructions
in this description.
⇨ Chapter 1 "Safety information", page 13
CAUTION!
A buffer battery is built into the device. Its purpose is to maintain data when the
device is switched off or if the voltage supply is interrupted. If the battery ap-
proaches the end of its operating life (approximately 7 years) then a battery
prealarm will indicate that status. When the battery is empty a battery alarm will
indicate that status. The battery must be exchanged before it is empty. The bat-
tery must be exchanged by the JUMO Service department! In this case, send
in the device!
CAUTION!
The touchscreen must not be operated with sharp or pointed objects, as these
could damage the protective film and the touchscreen.
Step Action
1 Switch on the supply voltage to the device and wait until it has boot-
ed up.
2 Select the operating language.
3 Log in as user "Master" or "Service" in order to have access to the
configuration in the device menu.
⇨ Chapter 8.2.1 "Log-on/Log-out", page 89
4 Set the date and time.
⇨ Chapter 9.1 "Date and time", page 127
5 Make the basic settings for the device.
⇨ Chapter 10.2 "Basic settings", page 134
6 Configure the analog and digital inputs you wish to use.
⇨ Chapter 10.5 "Analog inputs", page 139
⇨ Chapter 10.7 "Digital inputs of base unit and optional boards",
page 154
7 Check the hardware functions of the device.
⇨ Chapter 7.3 "Function test", page 73
8 Calibrate any analysis sensors connected to the device.
⇨ Chapter 12 "Calibration in general", page 205
9 The device is now ready for use. You can configure the displays
and functions of the device according to your needs.