8 Operation
8.1 Operating concept
This chapter explains how to use the functions at the operating level (e.g. con-
troller and data monitor) and access the menu structure for editing device set-
tings. The JUMO AQUIS touch P is operated by means of the touchscreen,
using either a finger or stylus with a soft, rounded plastic tip.
8.1.1 Passwords and user rights
The device has 4 users with factory-configured user names, passwords, and
user rights. The passwords can be changed in the device.
Chapter 8.2.1 "Log-on/Log-out", page 89
The JUMO PC setup program is needed to change user names and user rights.
"User list", page 313
The following table provides an overview of the factory-set user accounts.
Factory-set passwords
CAUTION!
The touchscreen must not be operated with sharp or pointed objects, as these
could damage the protective film and the touchscreen.
CAUTION!
Use only a soft cloth to clean the touchscreen. Commercially available cleaners
may contain substances that damage the protective film and the display.
NOTE!
Operation depends on the user's rights. Operating and setting options are re-
stricted, depending on the user who is logged in.
The "Master" and "Service" users have access to all menus and functions (de-
fault setting).
Chapter 8.1.1 "Passwords and user rights", page 75
Users Every-
one
User 2 User 1 Service Master
Factory-set password - 20 110 3000 9200