34
:ARM[:SEQuence[1]]
[:LAYer[1]]
:IMMediate
:COUNt <n>|
INFinite
:COUNt?
:SOURce HOLD|
IMMediate|
MANual|BUS|
TLINk|EXTernal
:SOURce?
:SIGNal
:TCONfigure
:DIRection
SOURce |
ACCeptor
:DIRection?
:ASYNchronous
:ILINe <n>
:ILINe?
:OLINe <n>
:OLINe?
Subsystem command path to
configure scan:
Path to program Arm Layer 1
(arm layer) of scan:
Loop around control source.
Program number of arms (1
to 9999, or INFinite).
Read arm count.
Select Arm Layer 1 control
source.
Read source.
Loop around control source.
Path to configure Triggers.
Enable (SOUR) or disable
(ACC) Bypass.
Read direction.
Path to configure asyn-
chronous Trigger Link:
Select input line (1 to 6).
Read input line.
Select output line (1 to
6).
Read output line.
Table 15: Trigger command summary (Cont.)
Command Description