Page 17
Find us at www.keysight.com
General Characteristics
Line power
Power supply (Mains supply voltage fluctuation is
not allowed to exceed ± 10%)
100 - 240 VAC
Power line frequency 50/60/400 Hz
Power consumption 45 watts
Environment
Operating environment Full accuracy - 0 to 55 C
Operating humidity Full accuracy to 80% R.H. at 40 °C non–condensing
(Humidity degrades linearly to 50%RH at 55 °C non-condensing)
Operating altitude up to 3,000 m
Storage temperature -40 to +70 C
Mechanical
Rack dimensions 212.6 mm x 88.5 mm x 348.3 mm
Bench dimensions 261.0 mm x 103.7 mm x 378.0 mm
Weight 4.1 kg
Regulatory
Safety EN 61010-1:2010 (3rd Edition)
ANSI/ISA-61010-1 (82.02.01) Third Edition
ANSI/UL 61010-1 Third Edition
CAN/CSA-C22.2 No. 61010-1 Third Edition
EN 61010-2-030:2010 (1st Edition)
ANSI/ISA-61010-2-030 (82.02.03) First Edition
ANSI/UL 61010-2-030 First Edition
CAN/CSA-C22.2 No. 61010-2-030 First Edition
Refer to Declaration of Conformity for current revisions
Pollution Degree 2
EMC IEC 61326
EN 61326
CISPR
ICES-001
AS/NZS 2064.1
Refer to Declaration of Conformity for current revisions
Acoustic noise - Nominal 35 dBA
Triggering conditions
External Delay
Jitter
Minimum Pulse width
Maximum rate
Voltmeter complete Polarity
Pulse width
Computer interfaces LXI (rev 1.4)
USB
LAN
Language SCPI-1999, IEEE-488
34970A/72A compatible
Front panel USB host port USB 2.0 - high speed mass storage (MSC) class device
Capability: Import/export instrument configuration files, save volatile readings and screen captures
Math Functions
Display
Real-time Clock
Software available IO Libraries: www.keysight.com/find/IOLibraries
BenchVue: www.keysight.com/find/benchvue