Chapter 4 Remote Interface Reference
Frequency Sweep Commands
4
Sweep Commands
FREQuency:STARt {<frequency>|MINimum|MAXimum}
FREQuency:STARt? [MINimum|MAXimum]
Set the start frequency (used in conjunction with the stop frequency).
Select from 1 mHz to 10 MHz
(limited to 100 kHz for a ramp waveform).
The default is 100 Hz. MIN = 1 mHz. MAX = 10 MHz. The :STAR? query
returns the start frequency in hertz.
• To sweep up in frequency, set the start frequency < stop frequency.
To sweep
down
in frequency, set the start frequency
>
stop frequency.
FREQuency:STOP {<frequency>|MINimum|MAXimum}
FREQuency:STOP? [MINimum|MAXimum]
Set the stop frequency (used in conjunction with the start frequency).
Select from 1 mHz to 10 MHz
(limited to 100 kHz for a ramp waveform).
The default is 1 kHz. MIN = 1 mHz. MAX = 10 MHz. The :STOP? query
returns the stop frequency in hertz.