Examples
The following example makes a frequency measurement on channel 1. The gate opens 5 events after a
negative edge on the Gate In/Out BNC, and, after .25 seconds, closes on the next positive edge on the
BNC.
CONF:FREQ 1.2E6
FREQ:GATE:SOUR ADV
GATE:STAR:SOUR EXT
GATE:STAR:SLOP NEG
GATE:STAR:DEL:SOUR EVEN
GATE:STAR:DEL:EVEN 5
GATE:STOP:SOUR EXT
GATE:STOP:SLOP POS
GATE:STOP:HOLD:SOUR TIME
GATE:STOP:HOLD:TIME .25
GATE:EXT:SOUR BNC
READ?
Typical Response: +1.21513398300000E+006
See Also
SENSe:FREQuency:GATE:SOURce
SENSe:TINTerval:GATE:SOURce
SENSe:TOTalize:GATE:SOURce
SENSe:GATE:STARt:SOURce
SENSe:GATE:STARt:SLOPe
SENSe:GATE:STOP:SOURce
SENSe:GATE:STOP:SLOPe