The following example returns 10 sets of 5 frequency measurements on channel 1, using a positive-going
external trigger to start each set of measurements.
CONF:FREQ 1.0E6,.001
SAMP:COUN 5
TRIG:COUN 10
TRIG:SOUR EXT
TRIG:SLOP POS
READ?
Typical Response: +1.00520000000000E+006, ... (50 measurements)
See Also
*RST
ABORt
SAMPle:COUNt
TRIGger:SOURce
TRIGger:COUNt
464 Keysight 53220A/53230A Programmer's Reference